Refurbished OEC 6600 C-arm
Refurbishing Process All OEC 6600 systems go through a multi-step refurbishing process that is the most extensive in the industry. Hand Surgery, Podiatry, Orthopedic, and Rehabilitation Services are just a few of the applications utilizing the a refurbished OEC 6600 C-arm. The CCD technology used in the OEC 6600 image system allows for crisp, clear imaging. For whatever the purpose, a refurbished OEC 6600 will deliver the results you can depend on.
Mechanical repairs/cosmetic maintenance
- All defective parts and covers are repaired or replaced
- Monitor cart is disassembled and repainted or retouched as needed
- X-ray tube removed and inspected for artifacts and internal integrity
- Image intensifier removed and inspected for any imperfections and grid integrity
- Camera removed, cleaned and tested
- Casters removed, cleaned and/or replaced
- All bearings are inspected, cleaned and re-lubricated, or replaced as needed
- Control panel display is tested and replaced if necessary
- All locks and brake assemblies are inspected, cleaned, and repaired or replaced as needed
- A complete set of operator and service manuals are provided
Re-assembly
- All components sent to C-arm reassembly area for rebuilding
- CRT monitors installed in monitor cart
- Monitor cart rebuilt, II, tube,camera and casters re-assembled
- When C-arm is completely rebuilt mechanically, system is sent for calibration and final testing
Testing/calibration
- Image intensifier and X-ray tube are tested for stability and balance
- Image intensifier is tested forresolution and gain to be withinOEM specifications
- X-ray tube radiation output is verified to be within OEMspecifications
- X-ray generator high voltage tested
- Maximum dose rate is set in compliance with government regulations
- CCD camera calibrated to OEM specifications
- X-ray beam is aligned for each field size
- C-arm is tested for current leak-age requirements
- All monitors are aligned, resized, centered and focused
- All monitors are adjusted for brightness and linearity
- Video system is optimized for gray scale and resolution
- All power supplies are tested and calibrated or replaced asnecessary
Specifications:
C-Arm Specifications:
- Free Space: 12”
- Depth: 14.5”
- SID: 14.9”
- Orbit Rotation: 110 degrees
- Rotation: +/- 200 degrees
- Wig-Wag: 180 degrees
- Vertical Travel: 24”
- Extension: Up to 55”
Mechanical Specs:
- Height: 64”
- Width: 27”
- Depth: 27”
- Footprint: 5 sq. ft.
Power Requirements:
- Input power: 60Hz
- 100-120V 15A
- 200-240V 8A