Reconditioned Philips Veradius C-arm

Reconditioning Process:
All equipment goes through a multi-step reconditioning process. Each unit is restored to a like-new status. The rotating x-ray tube allows for greater penetration and resolution for the lumbar area when doing needle localizations or other work involving the thicker anatomical areas of patients. For barium swallows a VCR or Digital Video Recorder can be utilized for better record keeping and post study analysis.

Mechanical repairs/cosmetic maintenance

  • All defective parts and covers are repaired or replaced
  • C-arm is completely disassembled and repainted
  • “C” is removed from main frame assembly
  • Horizontal arm removed, inspected and repaired as necessary
  • Vertical column removed and bearings removed inspected and replaced as required
  • X-ray tube removed and inspected for artifacts and internal integrity
  • Image intensifier removed and inspected for any imperfections and grid integrity
  • Camera removed
  • Wheels removed, cleaned and or replaced from main chassis
  • Chassis inspected for structural defects and repaired as necessary
  • Vertical lift assembly is tested for drift and adjusted/replaced if necessary
  • All bearings are inspected, cleaned and re-lubricated, or replaced as needed
  • Control panel display is tested and replaced if necessary
  • All wheels are inspected and lubricated and/or replaced as needed, with new cable pushers installed
  • All locks and brake assemblies are inspected, cleaned, and repaired or replaced as needed
  • C-arm is completely disassembled and repainted
  • Steering is tested and adjusted as required
  • Foot switch and hand switch provided on each unit
  • A complete set of operator and service manuals are provided

Re-assembly

  • All components sent to C-arm reassembly area for rebuilding
  • “C”, Image tube and camera, Vertical column, main control and power cable installed
  • New batteries are installed
  • New CRT monitors installed in monitor cart
  • Monitor cart rebuilt
  • When C-arm is completely rebuilt mechanically, system is sent for calibration and final testing

Testing/calibration

  • Image intensifier and X-ray tube are tested for stability and balance
  • Image intensifier is tested for resolution and gain to be within OEM specifications
  • X-ray tube bearings are tested for noise and coast time
  • X-ray tube filaments and stator windings are tested
  • X-ray tube radiation output is verified to be within OEM specifications
  • X-ray generator high voltage tested
  • Maximum dose rate is set in compliance with government regulations
  • CCD camera calibrated to OEM specifications
  • X-ray beam is aligned for each field size
  • C-arm is tested for current leakage requirements
  • All monitors are aligned, resized, centered and focused
  • All monitors are adjusted for brightness and linearity
  • Video system is optimized for gray scale and resolution
  • All power supplies are tested and calibrated or replaced as necessary